Models S310 & S310D Laser Power and Energy Meters Setup and Operating Procedures Serial Number___________________ P/N 1019

Strona 2

10ULTRA™ CALORIMETER SPECIFICATIONS: ModelUC150 UC150HD UC150HD40 UC150UVType of AbsorberSurface Surface Surface VolumeMaximum Beam Diameter25mm 25

Strona 3 - Table of Contents:

11QUICK SETUP: Note: For detailed instructions for each type of detector, refer to the Operating Procedures section. 1. Turn On the Meter: Note:

Strona 4 - ENVIRONMENTAL REQUIREMENTS:

125. Select a Range: On the first press of the RANGE button, the analog scale (if active) will disappear. Then briefly press the RANGE button each

Strona 5 - CAUTION: SPECULAR REFLECTION

132. Group Settings for Astral or Large Aperture Calorimeters: Note: The factory default Group for Astral or Large Aperture Calorimeters is Gr

Strona 6

14 To set the time delay between pulses use the count up (RANGE) and count down buttons (MODE) buttons to enter the time in seconds. Press the SELE

Strona 7

153. Group Settings for Vector Pyroelectric Detectors and Vector HR Pyroelectric Detectors: Note: The factory default Group for Vector Pyroel

Strona 8

16H. The "At" (attenuation) annunciator now appears in the display. This allows you to enter a value from 0.0001 to 9999 which will bec

Strona 9

17C. To select a Group, press the SELECT button when the Group you desire appears in the display (failure to press the SELECT button within 3 seco

Strona 10 - UNPACKING:

18OPERATING PROCEDURES: USING THE S310 WITH VECTOR PYROELECTRIC DETECTORS: Pyroelectric detector models P25, P50, SP25, SP50, P05, and P09 are coa

Strona 11 - 4. Connect a Detector:

193. To Zero the Analog Needle (does not apply to the S310D): Refer to Figure 3. The black slotted button located just below the display allows s

Strona 12 - 1. To Select a Group:

2Thank you for choosing a Scientech laser power and energy meter. Scientech, an ISO 9001 registered company, and our employees are pleased to prov

Strona 13

207. To Select a Mode: Available pyroelectric modes are: Energy (J), Avg. Energy (J AVG), Average Power (W AVG), and Volts (V). A. To Measure En

Strona 14

218. To Measure a Statistical Run of Energy Pulses: Note: Do not use AUTO range when making a statistical run. Note: Do not go from the Average

Strona 15 - Detectors:

2210. To Perform a Transfer Calibration: You can transfer a calibration from a calorimeter to a Vector pyroelectric detector using the Transfer Ca

Strona 16

23CORRECTING PYROELECTRIC DETECTOR OPERATING PARAMETERS FOR USE AT DIFFERENT WAVELENGTHS: NOTE: Due to variability in the manufacturing process t

Strona 17

242. To Turn the Analog Needle On or Off (does not apply to the S310D): To turn the analog meter on or off, press and keep holding the RANGE butto

Strona 18 - OPERATING PROCEDURES:

257. To Select a Mode: The modes available for the calorimeters are: Average Power (W AVG) or Energy (J). A. To Measure Average Power (W AVG):

Strona 19 - 6. To Select a Range:

26The statistics mode will collect data on a pulse population of up to 1000 pulses. At your prompting, the indicator will display the number of pul

Strona 20 - 7. To Select a Mode:

27 The following are viable external triggers. The following pulses will not trigger the S310. CORRECTING ASTRAL SERIES HD CALORIMETERS OPERAT

Strona 21 - 9. External Trigger:

28USING THE S310 WITH ULTRA SERIES DETECTORS: The Ultra detector selected needs to be the appropriate model for the planned laser measurements. P

Strona 22

29CORRECTING ULTRA SERIES HD CALORIMETERS OPERATING PARAMETERS FOR USE AT DIFFERENT WAVELENGTHS: NOTE: Due to variability in the manufacturing pr

Strona 23 - 1. To Turn On the Meter:

3Table of Contents: Detector Operating Parameters:...

Strona 24

301. Calorimeter Watts Mode: Compare voltage to the watt reading displayed on the S310 to determine V/W constant to use when looking at the ana

Strona 25

313. Pyroelectric Watts and Joules Modes: Compare peak voltage to energy (joules) or power (watts) reading displayed on the S310 to dete

Strona 26

32CALIBRATION OF ASTRAL CALORIMETERS USING ELECTRIC SUBSTITUTION HEATING: For Astral calorimeters the electric substitution heating option must be

Strona 27 - SE AT DIFFERENT WAVELENGTHS:

33CALIBRATION OF LARGE APERTURE CALORIMETERS USING ELECTRIC SUBSTITUTION HEATING: Electrical substitution heating is a standard feature of large ap

Strona 28

34where: V = voltage applied to the heater coil Rc = substitution heater resistance from step B C = Cal coefficient 360401 = 1.018 360801 =

Strona 29 - ANALOG OUTPUT:

35Operation of Astral and Large Aperture Calorimeters with a Digital Volt Meter: Note: Whenever a large aperture calorimeter is used without an i

Strona 30 - 1. Calorimeter Watts Mode:

36The error, in theory, is only dependent upon the value of ∑dVi, that is the cumulative random error of Vi. This number should approach zero if da

Strona 31 - 4. Ultra Calorimeter:

37Remote Interface Language: The remote interface language for the S310 is compatible with Std-IEEE488.2 and it also works with the RS-232 remote i

Strona 32 - HEATING:

384. Remote Interface Language Common Commands and Queries: Most commands and queries may be used with either the RS232 interface or the IEEE488 i

Strona 33 - SUBSTITUTION HEATING:

39RANGE_<dec num>|auto Sets the range to the decimal number supplied, or to the closest legal range allowed for the current detector and conf

Strona 34 - Pyroelectric Detectors:

43. To Zero the Analog Needle (does not apply to the S310D): ...

Strona 35

40 DUMP? If statistics have been collected from the front panel, or by the COL or COL? commands, this command returns all of the data values in the

Strona 36

41 *RCL_<dec num> This is the IEEE488.2 common command. It takes the single numeric argument [1-4], and restores the instrument configuratio

Strona 37 - 2. RS-232 Specifications:

425. IEEE488 Specific Commands: These commands may be used only from the IEEE488.2 interface. They are all members of the collection of the so-ca

Strona 38

43 RETURNED GOODS PROCEDURE: Should it become necessary to return any item to Scientech for any reason, please contact our Product Service Departme

Strona 39

44HD CALORIMETER AND PYROELECTRIC ABSORPTION VS. WAVELENGTH: Use this table for all HD calorimeters and HD pyroelectric detectors. Wavelength Absor

Strona 40

45 Wavelength Absorption Wavelength Absorption Wavelength Absorption Wavelength Absorptionµm% µm% µm% µm%1.245 91.78 1.510 91.70 1.775 90.84 2.016

Strona 41

46 Wavelength Absorption Wavelength Absorption Wavelength Absorption Wavelength Absorptionµm% µm% µm% µm%2.129 90.76 2.255 90.17 2.390 89.92 2.557

Strona 42 - LIMITED WARRANTY:

47 Wavelength Absorption Wavelength Absorption Wavelength Absorption Wavelength Absorptionµm% µm% µm% µm%2.855 89.26 3.233 88.79 3.725 88.11 4.394

Strona 43 - RETURNED GOODS PROCEDURE:

48Wavelength Absorption Wavelength Absorption Wavelength Absorption Wavelength Absorptionµm% µm% µm% µm%5.357 86.95 6.682 86.05 8.879 84.32 13.09 8

Strona 44

49P MODEL PYROELECTRIC DETECTOR ABSORPTION VS. WAVELENGTH: Use this table for standard and slim profile painted (P) model pyroelectric detectors. W

Strona 45

5VECTOR™ S310 INDICATOR SPECIFICATIONS: ModelS310 S310D4 Digit LCD with SelectableAnalog Meter MovementFull Scale Ranges with Astral 25mm Calorimet

Strona 46

50HF MODEL PYROELECTRIC DETECTOR ABSORPTION VS. WAVELENGTH: Use this table for standard and slim profile high frequency (HF) model pyroelectric det

Strona 47

51Wavelength Absorption Wavelength Absorption Wavelength Absorption Wavelength Absorptionµm% µm% µm% µm%2.067 60.27 2.257 60.06 2.487 59.79 2.768 5

Strona 48

52Wavelength Absorption Wavelength Absorption Wavelength Absorption Wavelength Absorptionµm% µm% µm% µm%3.120 59.47 3.576 59.37 4.188 58.92 5.052 5

Strona 49

53Wavelength Absorption Wavelength Absorption Wavelength Absorption Wavelength Absorptionµm% µm% µm% µm%6.365 57.97 7.870 57.74 10.31 60.33 14.93 4

Strona 50

54SCHEMATICS:

Strona 55

6VECTOR™ PYROELECTRIC DETECTOR SPECIFICATIONS: ModelP 25 PHF 25 PHD 25 PHDX 25 PHDX25UV SP 25 SPHF 25 SPHD 25Maximum Beam Diameter 25mm 25mm 25mm 7

Strona 56

7VECTOR  HR PYROELECTRIC DETECTOR SPECIFICATIONS: Model PHF 02 PHF 05 PHF 09 P 05 P 09Active Diameter 2 mm 5 mm 9 mm 5 mm 9 mmSpectral ResponseVol

Strona 57

8ASTRAL™ CALORIMETER SPECIFICATIONS: ModelAC2500 AC25HD ACX25HD AC2501 ACX2501 AC25UV AC2504Type of Absorber Surface Surface Surface Volume Volume

Strona 58

9LARGE APERTURE (100MM & 200MM) CALORIMETER SPECIFICATIONS: Model360401 380401 380402 384UV5 360801 380801 380802 384UV5Type of Absorber Surfac